Do bubbles present after adding red stop solution affect the final result?

Created by Alison Knapp, Modified on Tue, 12 Dec, 2023 at 3:24 PM by Pauline McCrystal

Since plate readers read through the liquid in the well from bottom to top, any bubbles present can affect the optical density readings and could potentially cause a failing r-value. Large bubbles in the centre of the well should be burst with a needle prior to reading. Very small bubbles around the edge of the wells should not have any impact on readings. 